Home/Recalls/FDA-Z-1767-2026
FDA DevicesClass II

SIGNA Premier systems

Published: April 8, 2026Recall ID: Z-1767-2026Category: devicesCountry: US

Reason for Recall / Hazard

Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.

Product Description & Identification

SIGNA Premier systems

Additional Source Details

FieldValue
CityFlorence
StateSC
Event id98579
Address 13001 W Radio Dr
Code infoUDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677
Postal code29501-7806
Report date20260408
Product typeDevices
Product quantity41
Reason for recallWhole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260227
Center classification date20260402

Overview

  • Recalling FirmGE Healthcare LLC
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of
Official Agency Alert