Home/Recalls/FDA-Z-2309-2026
FDA DevicesClass II

Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsu...

Published: June 10, 2026Recall ID: Z-2309-2026Category: devicesCountry: US

Reason for Recall / Hazard

3.0mm surgical knives labeled and packaged as 2.2mm surgical knives.

Product Description & Identification

Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife

Affected Products

Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife

Additional Source Details

FieldValue
CityChula Vista
StateCA
Event id98911
Address 11690 Brandywine Ave Ste C
Code infoUDI-DI 00848782027514; Lot Number FA22CHQ
Postal code91911-6072
Report date20260610
Product typeDevices
Product quantity1818 units
Reason for recall3.0mm surgical knives labeled and packaged as 2.2mm surgical knives.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260423
Initial firm notificationE-Mail
Center classification date20260604

Overview

  • Recalling FirmSSC- Surgical Specialties Corporation
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide - US Nationwide distribution in the states of Texas, Utah, Michigan and the countries of Turkey, Germany, Romania, Poland, United Kingdom.
Official Agency Alert