SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, So...

Published: June 17, 2026Recall ID: Z-2353-2026Category: devicesCountry: US

Reason for Recall / Hazard

Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.

Product Description & Identification

SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02

Additional Source Details

FieldValue
CityMogliano Veneto
StateN/A
Event id98924
Address 1Via Giotto 2
Address 2N/A
Code infoUDI-DI: 8033928120897; Lot Code: Model No SD LTM 64 PLUS; UDI-DI 8033928120897; Serial Numbers BAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19, BAA-0110/02-19, BAA-0111/02-19, BAA-0112/02-19, BAA-0113/02-19, BAA-0115/02-19, BAA-0116/02-19, BAA-0117/02-19, BAA-0118/02-19, BAA-0119/02-19, BAA-0121/02-19, BAA-0122/02-19, BAA-0142/03-19, BAA-0155/03-20, BAA-0156/03-20, BAA-0191/04-20, BAA-0192/04-20, BAA-0193/04-20, BAA-0194/04-20, BAA-0286/04-21, BAA-0287/04-21, BAA-0288/04-21, BAA-0289/04-21, BAA-0329/04-22
Postal codeN/A
Report date20260617
Product typeDevices
Product quantity207 units
Reason for recallNatus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260504
Initial firm notificationLetter
Center classification date20260605

Overview

  • Recalling FirmMicromed S.p.A.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide - US Nationwide distribution in the states of FL, PA.