Home/Recalls/FDA-Z-1713-2026
FDA DevicesClass I

NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,...

Published: April 15, 2026Recall ID: Z-1713-2026Category: devicesCountry: US

Reason for Recall / Hazard

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Product Description & Identification

NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.

Affected Products

NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.

Additional Source Details

FieldValue
CityNorthfield
StateIL
Event id98598
Address 13 Lakes Dr
Code info1. Medline SKU 70083007, UDI/DI each 10193489065855, UDI/DI case 30193489065859, Lot Numbers: 0000077029, 0000081537, 0000084335, 0000087347, 0000089563, 0000113792, 0000115753, 0000122956, 0000127785, 0000131986, 0000158748, 0000175535, 0000187926; 2. Medline SKU 70084007, UDI/DI each 10193489065893, UDI/DI case 30193489065897, Lot numbers: 0000078186, 0000081539, 0000084274, 0000086756, 0000089554, 0000094842, 0000096918, 0000105613, 0000109479, 0000111244, 0000114562, 0000115421, 0000117317, 0000118251, 0000120265, 0000120769, 0000121714, 0000123466, 0000124589, 0000131984, 0000131989, 0000138071, 0000140322, 0000142850, 0000147002, 0000152069, 0000162694, 0000173507, 0000174662, 0000174773, 0000176555, 0000178806, 0000189895; 3. Medline SKU 70084017, UDI/DI each 10193489065916, UDI/DI case 20193489065913, Lot Numbers: 0000084509, 0000086823, 0000095687, 0000165551, 0000165868, 0000166300; 4. Medline SKU 80083007, UDI/DI each 10193489066753, UDI/DI case 30193489066757, Lot Num... [TRUNCATED]
Postal code60093-2753
Report date20260415
Product typeDevices
Product quantity192690 units
Reason for recallMedline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260227
Initial firm notificationTwo or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Center classification date20260407

Overview

  • Recalling FirmMedline Industries, LP
  • StatusOngoing
  • Risk LevelClass I
  • DistributionWorldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).