FDA DevicesClass II
MEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 1...
Published: June 24, 2026Recall ID: Z-2481-2026Category: devicesCountry: US
Reason for Recall / Hazard
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Product Description & Identification
MEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407
Additional Source Details
| Field | Value |
|---|---|
| City | Northfield |
| State | IL |
| Event id | 98994 |
| Address 1 | 3 Lakes Dr |
| Code info | UDI/DI 10080196990740 (each), 40080196990741 (case); ALL LOTS |
| Postal code | 60093-2753 |
| Report date | 20260624 |
| Product type | Devices |
| Product quantity | 4304669 units |
| Reason for recall | During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260504 |
| Initial firm notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
| Center classification date | 20260617 |
Overview
- Recalling FirmMedline Industries, LP
- StatusOngoing
- Risk LevelClass II
- DistributionWorldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.