Home/Recalls/FDA-Z-2484-2026
FDA DevicesClass II

MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr...

Published: June 24, 2026Recall ID: Z-2484-2026Category: devicesCountry: US

Reason for Recall / Hazard

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Product Description & Identification

MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C

Additional Source Details

FieldValue
CityNorthfield
StateIL
Event id98994
Address 13 Lakes Dr
Code infoUDI/DI 10197344149001 (each), 20197344149008 (case); ALL LOTS
Postal code60093-2753
Report date20260624
Product typeDevices
Product quantity10980 units
Reason for recallDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260504
Initial firm notificationTwo or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Center classification date20260617

Overview

  • Recalling FirmMedline Industries, LP
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Official Agency Alert