FDA DevicesClass II
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Published: April 22, 2026Recall ID: Z-1823-2026Category: devicesCountry: US
Reason for Recall / Hazard
The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
Product Description & Identification
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Affected Products
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Additional Source Details
| Field | Value |
|---|---|
| City | Sudbury |
| State | MA |
| Event id | 98619 |
| Address 1 | 142 North Rd Ste G |
| Address 2 | N/A |
| Code info | Model Number: G10-01; UDI-DI: 00859910007032; Lot/Serial Number: 23C0113, 22J0101, 20A0202, 23F0101, 22F0113, 22L0100, 26B0203, 26B0200, 26B0204, 26B0202, 26B0201, 25L0302, 22E0204, 22J0102, 24H0101, 24H0102, 23K0201, 20A0203, 22K0306, 20A0201, 25J0200, 23C0110, 22K0305, 25E0100, 22F0110, 22K0304, 22K0302, 24B0107; OUS Lot numbers: 21G1900, 21G1901, 21G1902, 21G1903, 21G1904, 21H1101, 21H1102, 21H1103, 21H1104, 21H1105, 22C2300, 22C2301, 22C3000, 22D2600, 22D2700, 22D2701, 22D2702, 22D2705, 22E0500, 22E0501, 22F2700, 22F2701, 22F2702, 22F2703, 22F2704, 22G2600, 22G2601, 22G2700, 22H2900, 22H2901, 24B2100, 24B2101, 24B2102, 24B2103, 24B2104, 24C1301, 24C1302, 24C1303, 24C1304, 24C1305, 24D1504, 24D1505, 24D1506, 24D1507, 24D1508, 24E0800, 24E0801, 24E0802, 24E0803, 24E0804, 24F2408, 24F2409, 24F2410, 24F2411, 24F2412, 24G2205, 24G2206, 24G2207, 24G2208, 24G2209, 24H1400, 24H1401, 24H1402, 24H1403, 24H1404, 24M1000, 22L3000, 22L3001, 23A1904, PDA23-001, 23C2202, 23C2203, 23E0900, 23E0901... [TRUNCATED] |
| Postal code | 01776-1142 |
| Report date | 20260422 |
| Product type | Devices |
| Product quantity | 184 units (28 US, 156 OUS) |
| Reason for recall | The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260323 |
| Initial firm notification | Letter |
| Center classification date | 20260416 |
Overview
- Recalling FirmGentuity, LLC
- StatusOngoing
- Risk LevelClass II
- DistributionDomestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;