Reason for Recall / Hazard

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Product Description & Identification

GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Additional Source Details

FieldValue
CityWaukesha
StateWI
Event id98938
Address 13114 N Grandview Blvd
Code infoSerial #/Sales Order #: SS920500004SA, SS920500005SA, SS920500003SA, SS920500006SA, SS921120011SA, SS921120019SA, SS921120015SA.
Postal code53188-1615
Report date20260603
Product typeDevices
Product quantity7 units
Reason for recallGE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260424
Initial firm notificationLetter
Center classification date20260528

Overview

  • Recalling FirmGE Medical Systems Information Technologies Inc
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide distribution.