Description/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC ...
Reason for Recall / Hazard
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Product Description & Identification
Description/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-12703-PSPH1; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-WMC2; MULTILUMEN PI CVC KIT: 3L 7 FR X 20CM/ASK-15703-LMDT1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20CM/ASK-15703-LMDTG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-15703-PVH; 1L CVC KIT: 16 GA X 8 IN (20CM)/ASK-24301-JHH1; CVC KIT: 16 GA X 16 CM/CDC-24306-1A; CVC KIT: 3-L 7 FR X 30 CM ANTIMICROBIAL/CDC-24703-X1A; LBCVC KIT: 3L 12 FR X 20 CM AGB/CDC-25123-1A; CVC KIT: 2L 8 FR X 6IN (16 CM)/ASK-42802-ECL3; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-ACH; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-BJC1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16 CM/ASK-42703-DU1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-EMC; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6IN/ASK-42703-LMDT1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-LMDTG1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-MCY1; MULTILUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-NS1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PAGP1; PI CVC KIT: 3L 7 FR X 6 IN (16 CM)/ASK-42703-PAM1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PAMC2; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PAU1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16/ASK-42703-PB1; PI CVC KIT: 3L 7 FR X 16CM AGB+/ASK-42703-PBSWH2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16 IN/ASK-42703-PCAM1; MULTILUMEN PI CVC KIT: 3L 7 FR X 16 CM/ASK-42703-PCSJ2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PECL2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PGM4; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PHHN1; MULTI LUMEN CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PLC; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMEM; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMHC1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-POHU2; MULTILUMEN PI CVC KIT: 3L 7 FR X 6/ASK-42703-POSU2; MULTILUMEN CVC KIT 7 FR X 6 IN (16 CM)/ASK-42703-PTJH; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PUCH2; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-PUCI1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PUHC2; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PUMH; MULTI-LUMEN PI CVC KIT: 3-L 7 FR/ASK-42703-PUPM1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-SHUF2; MULTI LUMEN PI CVC KIT: 3L 7 FR/ASK-42703-SL1; CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-42703-UMC; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-WHH; CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-CHM; CVC KIT: 2L 8 FR X 16 CM/ASK-42802-LMDT1; PI CATHETER 2L: 8FR X 16CM AGB+/ASK-42802-MCY1; PI CVC KIT: 2L 8FR X 16CM AGB+/ASK-42802-PBSWH2; MULTILUMEN CVC KIT: 2L 8 FR X 6/ASK-42802-POSU2; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PUNCC2; PI CVC KIT: 2-L 8 FR/ASK-42802-PUPM1; PI CVC KIT: 2L 8 FR/ASK-42802-TGH1; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-BJC1; CVC KIT: 4L 8.5 FR X (16CM) AGB+/ASK-42854-PBMH; PI CVC KIT: 4L 8.5FR X 16CM AGB+/ASK-42854-PMM; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-PUCH2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-ACH; 3L CVC KIT: 7 FR X 20 CM AGB+/ASK-45703-ARM; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-BJC1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-BLH; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-DU1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-HHC4; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-LMDT1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-LMDTG1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-MCY1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-NO; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-NS1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PADV1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PAGP1; MULTI LUMEN CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PAMC1; MULTI LUMEN CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PAU1; PI CVC KIT: 3L 7FR X 20CM AGB+/ASK-45703-PBSWH2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PCAM1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PCSJ2; PI CVC KIT: 3L 7 FR X 8 IN (20 CM) AGB+/ASK-45703-PDM1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PECL2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/A
Additional Source Details
| Field | Value |
|---|---|
| City | Morrisville |
| State | NC |
| Event id | 98793 |
| Address 1 | 3015 Carrington Mill Blvd |
| Code info | REF(Material)/UDI-DI/Lot(Batch): ASK-04001-JHH/10801902193275/33F24K0748, 33F26A1669, 33F25G0558; ASK-04001-LMDTG/10801902193275, 10801902218442/33F26A0405, 33F24E0333; ASK-12703-PSPH1/10801902214338, 10801902193275, 10801902214338/33F24F0787, 33F24L0304, 33F25A0242, 33F25C0268, 33F25D0470, 33F25E0537, 33F25E0854, 33F25J0365, 33F25M0247, 33F26A1555, 33F26B0855, 33F26C0742, 33F24K0672; ASK-12703-WMC2/10801902193275/33F24K0493, 33F24L0305, 33F25B0156, 33F25E0593, 33F25H0483, 33F25J0367, 33F25M0218, 33F24K0545; ASK-15703-LMDT1/10801902193275/33F24J0164, 33F24L0307, 33F25B0192, 33F25C0230; ASK-15703-LMDTG1/10801902193275/33F25D0537, 33F25E0670, 33F25G0295, 33F25J0146, 33F25K0093, 33F25M0138, 33F26B0131, 33F25E0312; ASK-15703-PVH/10801902133226, 10801902193275, 10801902210217/33F24D0809, 33F24F0669, 33F24K0462, 33F24L0226, 33F25A0167, 33F25B0193, 33F25B0547, 33F25D0265, 33F25D0653, 33F25G0169, 33F25G0404, 33F25J0369, 33F26A1459, 33F24M0270; ASK-24301-JHH1/10801902193275, 108019022148... [TRUNCATED] |
| Postal code | 27560-5437 |
| Report date | 20260624 |
| Product type | Devices |
| Product quantity | 1,061,069 |
| Reason for recall | Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260421 |
| Initial firm notification | Letter |
| Center classification date | 20260612 |
Overview
- Recalling FirmARROW INTERNATIONAL, LLC
- StatusOngoing
- Risk LevelClass I
- DistributionUS Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.