Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE ...
Reason for Recall / Hazard
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Product Description & Identification
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Affected Products
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Additional Source Details
| Field | Value |
|---|---|
| City | Center Valley |
| State | PA |
| Event id | 98422 |
| Address 1 | 3500 Corporate Pkwy |
| Code info | Laser System: TFL-SLS; UDI: 00821925044135; Component Part: TFL-CSLU; UDI: 00821925044593; Serial #: MDUF190114, MDUF220073, MDUF220331, MDUF220353, MDUF220354, MDUF220356, MDUF220411, MDUF220412, MDUF220413, MDUF220414, MDUF220427, MDUF220451, MDUF220453, MDUF220454, MDUF220455, MDUF220456, MDUF230004, MDUF230005, MDUF230006, MDUF230007, MDUF230076, MDUF230140, MDUF230141, MDUF230163, MDUF230164, MDUF230166, MDUF230167, MDUF230234, MDUF230283, MDUF230284, MDUF230331, MDUF230332, MDUF230361, MDUF230362, MDUF230363, MDUF230364, MDUF230365, MDUF230366, MDUF230411, MDUF230412, MDUF230413; |
| Postal code | 18034-8229 |
| Report date | 20260408 |
| Product type | Devices |
| Product quantity | 32 units |
| Reason for recall | Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. |
| Voluntary mandated | Voluntary: Firm initiated |
| Recall initiation date | 20260227 |
| Center classification date | 20260330 |
Overview
- Recalling FirmOlympus Corporation of the Americas
- StatusOngoing
- Risk LevelClass II
- DistributionWorldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.