Reason for Recall / Hazard

Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

Product Description & Identification

Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No

Affected Products

Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No

Additional Source Details

FieldValue
CityCenter Valley
StatePA
Event id98422
Address 13500 Corporate Pkwy
Code infoLaser system: TFL-PLS; UDI: 00821925044111; Component Part: TFL-CPLU; UDI: 00821925044586; Serial #: MDUF220005, MDUF220005, MDUF220437, MDUF220437, MDUF220446, MDUF220446, MDUF190072, MDUF200324, MDUF210030, MDUF210204, MDUF210379, MDUF220001, MDUF220002, MDUF220003, MDUF220007, MDUF220008, MDUF220009, MDUF220010, MDUF220011, MDUF220012, MDUF220013, MDUF220014, MDUF220015, MDUF220016, MDUF220017, MDUF220018, MDUF220019, MDUF220020, MDUF220021, MDUF220022, MDUF220023, MDUF220024, MDUF220025, MDUF220026, MDUF220027, MDUF220028, MDUF220029, MDUF220030, MDUF220031, MDUF220032, MDUF220033, MDUF220034, MDUF220035, MDUF220036, MDUF220037, MDUF220038, MDUF220039, MDUF220040, MDUF220042, MDUF220043, MDUF220044, MDUF220045, MDUF220046, MDUF220047, MDUF220048, MDUF220049, MDUF220050, MDUF220051, MDUF220052, MDUF220053, MDUF220054, MDUF220055, MDUF220056, MDUF220057, MDUF220058, MDUF220060, MDUF220061, MDUF220062, MDUF220063, MDUF220064, MDUF220065, MDUF220066, MDUF220067, MDUF22006... [TRUNCATED]
Postal code18034-8229
Report date20260408
Product typeDevices
Product quantity408 units
Reason for recallFollowing a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260227
Center classification date20260330

Overview

  • Recalling FirmOlympus Corporation of the Americas
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.