FDA DevicesClass II

Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003...

Published: June 3, 2026Recall ID: Z-2259-2026Category: devicesCountry: US

Reason for Recall / Hazard

It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

Product Description & Identification

Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039);

Additional Source Details

FieldValue
CityBest
Event id98826
Address 1Veenpluis 6
Code info1. Allura Xper FD10; Model: 722003; UDI-DI: N/A; Serial Numbers: 445, 1555, 1114, 1529, 472, 1349, 1594, 1311, 1472, 1187, 1700, 1031, 912, 913, 940, 285, 1637, 5, 797, 639, 452, 1585, 535, 1338, 1641, 1916, 614, 605, 492, 1612, 505, 239, 1216, 1169, 1118, 115, 519, 356, 715, 423, 1399, 789, 1509, 1016, 875, 1923, 313, 314, 1919, 1458, 369, 1176, 1230, 1925, 339, 1325, 1633, 596, 1784, 464, 796, 1928, 1025, 292, 1669, 1697, 708, 1, 1760, 1452, 346, 961, 1172, 354, 416, 363, 582, 161, 569, 922; 2. Allura Xper FD10; Model: 722010; UDI: (01)00884838059030(21)1123, (01)00884838059030(21)932, (01)00884838059030(21)239, (01)00884838059030(21)360, (01)00884838059030(21)500, (01)00884838059030(21)794, (01)00884838059030(21)939, (01)00884838059030(21)921, (01)00884838059030(21)286, (01)00884838059030(21)1440, (01)00884838059030(21)1144, (01)00884838059030(21)521, (01)00884838059030(21)33, (01)00884838059030(21)107, (01)00884838059030(21)1025, (01)00884838059030(21)668, (01)00884838059030(21)656... [TRUNCATED]
Postal codeN/A
Report date20260603
Product typeDevices
Reason for recallIt was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260430
Initial firm notificationLetter
Center classification date20260528

Overview

  • Recalling FirmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.