Home/Recalls/FDA-Z-2318-2026
FDA DevicesClass II

Allia Moveo angiographic X-ray system

Published: June 17, 2026Recall ID: Z-2318-2026Category: devicesCountry: US

Reason for Recall / Hazard

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Product Description & Identification

Allia Moveo angiographic X-ray system

Additional Source Details

FieldValue
CityWaukesha
StateWI
Event id98880
Address 13000 N Grandview Blvd
Address 2N/A
Code infoUDI-DI: 00198953052488; Serial Numbers: C4-25-001, C4-25-002, C4-26-001, C4-26-002, C4-26-004, C4-26-005
Postal code53188-1615
Report date20260617
Product typeDevices
Product quantity3 units
Reason for recallGE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260420
Initial firm notificationLetter
Center classification date20260605

Overview

  • Recalling FirmGE Medical Systems, LLC
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide - US Nationwide distribution.
Official Agency Alert