Home/Recalls/FDA-Z-1874-2026
FDA DevicesClass II

AccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers; ACC...

Published: April 29, 2026Recall ID: Z-1874-2026Category: devicesCountry: US

Reason for Recall / Hazard

The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.

Product Description & Identification

AccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 21 R/O 1CM; Catalog No.: 13709704; Product/UPN No.: H787137097045 (Box), H787137097040 (Pouch); Box Quantity: 5 pouches;

Affected Products

AccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 21 R/O 1CM; Catalog No.: 13709704; Product/UPN No.: H787137097045 (Box), H787137097040 (Pouch); Box Quantity: 5 pouches;

Additional Source Details

FieldValue
CityQueensbury
StateNY
Event id98463
Address 1603 Queensbury Ave
Address 2N/A
Code infoACCU-VU OF 4F X 70CM 035 NB 6SH 21 R/O 1CM; Catalog No.: 13709704; Product/UPN No.: H787137097045 (Box), H787137097040 (Pouch); UDI-DI: 25051684006030(Box), 15051684006033(Pouch); Lot No.: A3025039, A3225033;
Postal code12804-7619
Report date20260429
Product typeDevices
Product quantity500 units (100 boxes)
Reason for recallThe affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
Voluntary mandatedVoluntary: Firm initiated
Recall initiation date20260303
Initial firm notificationLetter
Center classification date20260417

Overview

  • Recalling FirmAngiodynamics, Inc.
  • StatusOngoing
  • Risk LevelClass II
  • DistributionWorldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, China, Chile, Curacao, Czech Republic, Germany, Ireland, Israel, Italy, Lebanon, Netherlands, New Zealand, Panama, Slovenia, South Africa, Spain, Sweden, Thailand.
Official Agency Alert