Medical Device
FDA DevicesClass II

Brand Name: DONJOY Product Name: ICEMAN, W/COLDPAD, NS, RH Model/Catalog Numbe...

Due to broken connector which may result in a temporary interruption of prescribed cold therapy.

May 20, 2026DJO, LLC
Medical Device
FDA DevicesClass II

Brand Name: DONJOY Product Name: ICEMAN CLASSIC CUBE Model/Catalog Number: 11-...

Due to broken connector which may result in a temporary interruption of prescribed cold therapy.

May 20, 2026DJO, LLC
Medical Device
FDA DevicesClass I

Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS...

16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.

May 20, 2026ARROW INTERNATIONAL, LLC
Medical Device
FDA DevicesClass II

Daig Livewire Steerable, Product Number 401923; REPROCESSED ELECTROPHYSIOLOGY CA...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass II

Daig Livewire Steerable, Product Number 401908; REPROCESSED ELECTROPHYSIOLOGY CA...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass II

Thermedx FluidSmart Urology Tube Set REF LL0006 UDI-DI code: 00860936000365 ...

Due to nonconforming products being inadvertently distributed.

May 20, 2026Stryker Corporation
Medical Device
FDA DevicesClass I

Rolls of label stock used with the Omnicell i.v.Station. 1. Part Number 258...

Potential for mislabeled syringe produced by the i.v.STATION device.

May 20, 2026Omnicell, Inc.
Medical Device
FDA DevicesClass II

BARD Dynamic XT Steerable, Product Number 201103, REPROCESSED ELECTROPHYSIOLOGY ...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass I

Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-V...

16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.

May 20, 2026ARROW INTERNATIONAL, LLC
Medical Device
FDA DevicesClass I

NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SF...

16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.

May 20, 2026ARROW INTERNATIONAL, LLC
Medical Device
FDA DevicesClass II

Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27

GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

May 20, 2026GE Medical Systems, LLC
Medical Device
FDA DevicesClass II

On-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33

Heart valves were released and distributed before all required testing was complete, so it cannot be confirmed that they meet all release specifications.

May 20, 2026On-X Life Technologies, Inc.
Medical Device
FDA DevicesClass II

SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-...

Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

May 20, 2026Hologic, Inc
Medical Device
FDA DevicesClass II

LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWork...

Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events

May 20, 2026MICROVENTION INC.
Medical Device
FDA DevicesClass II

Daig Livewire Steerable, Product Number 401652; REPROCESSED ELECTROPHYSIOLOGY CA...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass II

Diowave Laser System, REF: Diowave 250W

Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthritis, and to temporarily increase local blood circulation, operating at a maximum average power output of 70W, which is beyond the FDA cleared 60W.

May 20, 2026Technological Medical Advancements LLC
Medical Device
FDA DevicesClass II

Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,RH Model/Catalog Numbe...

Due to broken connector which may result in a temporary interruption of prescribed cold therapy.

May 20, 2026DJO, LLC
Medical Device
FDA DevicesClass II

Daig Livewire Steerable, Product Number 401575; REPROCESSED ELECTROPHYSIOLOGY CA...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass II

BARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOG...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass II

Daig Livewire Steerable BDB, Product Number 401918, REPROCESSED ELECTROPHYSIOLOG...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass II

Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20

GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

May 20, 2026GE Medical Systems, LLC
Medical Device
FDA DevicesClass II

BARD EP XT Decapolar Steerable, Product Number 201007; REPROCESSED ELECTROPHYSIO...

Incomplete seals on sterile product

May 20, 2026Stryker Sustainability Solutions
Medical Device
FDA DevicesClass I

Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038200; Cardio...

Medline has identified the presence of particulate within the fluid path of the Manifolds.

May 20, 2026Medline Industries, LP
Medical Device
FDA DevicesClass II

Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,EH Model/Catalog Numbe...

Due to broken connector which may result in a temporary interruption of prescribed cold therapy.

May 20, 2026DJO, LLC